Purge-Gas-Generators (FTIR)
Generating purge gas out of compressed air The new Purge-Gas Generators of the PG series are produced specially for the use with FT-IR Spectrometers to provide a purified purge gas and air bearing gas from compressed air. The generator is delivered ready to use with coalescing prefilters and automatic drains. First oil-, water- and dust particles are removed from compressed air. CO2 removal is achieved by pressure swing adsoption (PSA) technology to reduce the concentration of these compounds to CO2 < 1 ppm, H2O < 1 ppm. The molecular sieve is regenerated automatically by removing CO2 and H2O. The process is being repeated continuously in time and nearly maintenance-free. To protect the whole system it is recommended to change once per year the prefilter. All systems are designed to operate continuously 24 hours/day, 7 days/week. Compressed air supply All Purge-Gas Generators are available with a external low noise and oil-less compressor for permanent use (<55 dba). Max. air inlet temperature + 30 °C. Air should be prefiltered and oil free. To guarantee a longlasting maintenance-free use please use air with the following standards: ISO 8573.1, class 1.2.1. Oilcontent < 0,01 mg/m3, water content < - 40 °C dew point, particles < 0,1 µm. In case of dirty compressed air with maybe high rate of oil and water special coalscing-filters can be delivered. | Benefits Using a Purge-Gas Generator of the PG series the FT-IR chamber will be removed much quicker from CO2 and water than using nitrogen. The user gets clearer backgrounds in a shorter period of time. The small and space saving systems use your in house compressed air supply. Therefore dangerous and expensive gas cylinders are eliminated from the laboratory. |
Technical data
Type | PG 14 L | PG 28 L | PG 85 L |
Max. dry air flowrate for specified dew point [Nl/min] | 14 | 28 | 85 |
Max. flowrates higher than 85 Nl/min. are also available | |||
Max. CO2- and H20-content at the outlet [ppm] | < 1 ppm < 0,1 ppm | ||
Connections (Swagelok®) [mm] | 6 | ||
Recommended inletpressure [bar] | 5-10 | ||
Electrical requirement | 230 V/50 Hz | ||
Power consumption [Wh] | < 10 | < 20 | < 30 |
Performance max. [W] | 25 | 30 | 50 |
Dimensions (W x H x D) [mm] | 180 x 370 x 300 | 180 x 370 x 300 | 400 x 600 x 400 |
Weight [kg] | 8,0 | 12,5 | 28,0 |
Download Specification |